AMI Turbiwell 7027
High precision, non-contact nephelometer based on ISO 7027 (EN 27027, DIN 38404) for the continuous measurement of turbidity in pure water, potable water, surface water treatment and effluent.
- Range: 0.000 to 200.0 FNU with a precision of ±(0.003 FNU + 1 % of reading).
- Turbidimeter with sample chamber and valve for chamber drain.
- Optional sample degasser for the prevention of interfering bubbles formation in the sample.
- High precision secondary standards for quality assurance.

Specifications
|
Measuring Range:
0 - 200.0 FNU/NTU
Precision:
± (0.003 FNU + 1% of reading)
Accuracy (based on formazine primary standard):
- Range 0 - 40 FNU ± (0.01 FNU + 2% of reading)
- Range > 40 FNU: ± 5% of reading
Sample Conditions:
Flow rate: Approx. 20 - 60 l/h
Temperature:
up to 45°C
Power Supply:
- 00 - 240 VAC (±10%)
- 10 - 36 VD